Stay FreshCommercial

A shelf freeze dryer for the research lab

Stay Fresh freeze dryers are food machines with lab-grade bones: a programmable nine-step shelf temperature curve from −50 to 180 °F, a KF25 vacuum port, a temperature probe input, and a release valve you can feed with dry nitrogen. They suit R&D, teaching and sample work at a fraction of a lab lyophilizer’s price. They are not for GMP production, and this page says exactly why.

By the Stay Fresh engineering team. Updated November 12, 2026.

Glass vials on a shelf inside a Stay Fresh Mega freeze dryer
Vials on a shelf in a Stay Fresh Mega.

What it does well

  • Cycle development. Advanced mode runs a preset shelf temperature curve with nine slots, each a target shelf temperature and a hold time in minutes, so you can program a freeze hold, an annealing step, staged primary drying and a secondary drying step.
  • Aqueous samples, reagents, buffers and media, plant and food-science samples, and archived material that has to come out dry and stay that way.
  • Teaching labs that need a real shelf freeze dryer students can program, at a price a department budget absorbs.
  • Inert backfill. Breaking the vacuum with dry nitrogen through the dedicated release valve, rather than room air, keeps finished cakes out of humid air while you cap them.
  • Instrumenting a batch. RTD shelf sensing plus a plug-in probe for the product (2-pin on the Large and Mega in standard, needle or pad form; 3-pin on the Standard), and a KF25 flange for connecting gauges or a different pump.

What it is not

Stated plainly, so nobody finds out late. These machines have no in-chamber stoppering, no clean-in-place or steam-in-place, no sterile boundary, no validated control system and no audit trail meeting 21 CFR Part 11. Target vacuum is one setting for the whole run, not per step. They are not suitable for making sterile product for human use. The manual also warns that misusing the advanced modes voids the vacuum pump warranty, a fair warning on a pump that isn’t meant to run cold and wet for days.

They are built for aqueous samples and food. Talk to us before running solvents, acids or anything corrosive, which can damage the pump and seals, and don’t run infectious material: there is no biological containment.

Choosing a chamber for lab work

Two numbers decide it: the room between shelves, and how cold the trap runs. The trap temperature sets the lowest chamber pressure a load can hold, because the chamber can’t go below the vapor pressure of the ice already on the trap.

ChamberShelf spacingCold trapIce vapor pressure at the trapTray areaPrice
5-Shelf Commercial1.6 in (40.6 mm)−40 °F (−40 °C)about 97 mTorr800 sq in$3,360
Large1.6 in (40.6 mm)−40 °F (−40 °C)about 97 mTorr1,283 sq in$4,335
Mega2 in (50.8 mm)−35 °F (−37.2 °C)about 131 mTorr1,733 sq in$5,355

For vials, the Mega. Its 2 in shelf pitch fits 2R and 6R vials with a lyophilization stopper half-seated; 4R, 8R and 10R are marginal, and 20R and up need a shelf removed. Shelf spacing is measured shelf to shelf, so measure the clear height in your machine against your own vial stock before you plan around a marginal size.

For the most collapse-sensitive formulations, the Standard or Large. Sucrose has a Tg′ of about −33 °C. Holding the sublimation front below that needs the chamber near 200 mTorr, which all three can do, but the −40 °F trap leaves a little more margin than the Mega’s −35 °F. Trehalose, at about −29 °C, is comfortable on any of them.

On an empty shelf, these machines typically pull below 200 mTorr within about 15 minutes. Under load the pressure you can hold is higher, because the vapor the product releases is exactly what the pump and trap have to keep up with; measure where your loaded chamber settles. The target vacuum figures in our food recipes are food settings, not machine limits.

Price, in context

Lab freeze dryers come in different forms, so compare on what you need to dry. A shelf freeze dryer with a programmable shelf curve and its pump included costs $3,360 to $5,355 here. For comparison, a Labconco FreeZone 2.5 L −50 °C benchtop unit sells for about $8,500 to $12,100 from dealers, and the 8 L −50 °C model at $20,040 to $21,390 at Fisher Scientific, whose listing notes that a drying accessory, vacuum pump, glassware and adapters are needed on top (prices seen November 2026). Those instruments bring certifications, data logging and validation support that ours don’t. If your work needs them, buy one; if it doesn’t, a shelf freeze dryer gets you drying for a fraction of the price.

A starting cycle

Our guide to freeze-drying vials in the Mega works through vial heat transfer, fill depth, the edge-vial effect, annealing mannitol, and a nine-slot starting curve for a mannitol-containing formulation at 200 to 300 mTorr, about 41 hours from freeze to secondary drying. It is a starting point, not a validated recipe.

Three habits matter more than any setting: put a probe in a center vial and treat it as an indicator rather than a measurement of the batch; ring the vials you care about with water-filled dummies to absorb the extra radiant heat at the edges; and confirm residual moisture by Karl Fischer titration if anything downstream depends on it.

Buying for a lab

We send written quotes that a purchasing department can work from, for one machine or several, in 110 V or 220–240 V. Request a quote and mention what you’ll be drying; technical questions are answered free by phone, email or video call, before and after you buy.

Quick answers

Can it run a lyophilization cycle with annealing and secondary drying?

Yes, through the nine-step preset temperature curve in Advanced mode. Each step sets a shelf temperature and a hold time. There’s no per-step vacuum control: target vacuum is one setting for the whole run.

Can it stopper vials?

No. There’s no in-chamber stoppering. Backfill with dry nitrogen through the release valve, then cap each vial as you take it out, in the driest space you have or a nitrogen-purged glove bag.

How cold does the trap get?

−40 °F (−40 °C) on the Standard and Large, −35 °F (−37.2 °C) on the Mega. Lab benchtop units often run −50 °C to −84 °C or colder, which matters for low-collapse-temperature formulations and solvents.

Is it suitable for GMP work?

No. There’s no validated control system, audit trail, sterile boundary, CIP/SIP or stoppering. It is for research, development and teaching.